32L Lab Use Pure Nitrogen Gas Generator Suitable for HPLC ICPMS Introduce

Application of 32L nitrogen gas generator:

1.Nitrogen supply for Termovap Sample Concentrator

2.Nitrogen supply for LCMS

3.Nitrogen supply for ICPMS

4.supply nitrogen for HPLC

5.Other pure nitrogen application field

Detailed technical specification of PSA32L nitrogen gas generator:

2.1 Compressed air consumption

The generator consumes a maximum of 115 slpm of compressed air to generate the required N2 flow. Air source is not

included in this product.

2.2. Air supply pressure

Minimum air supply pressure shall be externally regulated to 600 kPa.

2.3. Air quality

This Generator is provided with one full set of FESTO filters, the air supplied to the Generator still need to be dry and

oil-free.

2.4. N2 output

2.4.1. Quantity and purity of N2 generate:N2 flow rate scope 0-32L/min

2.4.2. Delivery pressure: the generator delivers N2 at a pressure greater than 450 kPa.

2.4.3. O2 contamination level: O2 level in the N2 stream is not exceed 0.5%, and does not vary from peak to trough by

more than 5% of the absolute concentration.

2.5 Dimensions

With package, the Generator size: 850 mm (Length)x 920 mm (width)x 1100 mm(height).

2.6 Weight

With package, the Generator weight is not more than 155 kg.

2.7 Wheels

The Generator is fitted with 2 fixed wheels at the rear and 2 swivel wheels with brakes at the front

2.12. Start up

2.12.1. Start up time

Start up times means the time from start up to producing qualified nitrogen.

The time to complete start up is less than 15 minutes.

2.13. Gas connections

The two external gas connections on the Generator, Air In and N2 Out, are constructed from brass and are to suit ¼”

(6.35 mm) OD tubing. There are labeling in English next to these connections, “Air In” and “N2 Out”, as part of the

rating label.

2.14. Environmental requirements

2.14.1. Operating temperatures

The Generator is normally operated within performance specification over the temperature range:

10ºC - 35ºC from 0 – 3000 m, and 10ºC - 25ºC from 3000 – 4500 m

The Generator can operate safely over the temperature range:

5ºC - 35ºC from 0 – 3000 m, and 5ºC - 25ºC from 3000 – 4500 m

2.14.2. Noise level requirements

Noise level from the Generator is less than 72dBa at 1m distance. This is achieved by fitting silencers on the exhaust

vents.

Please no hesitate to contact us if any question