Class ll type A2 biological safety cabinet that protects the user and lab environment from the biohazard factor and prevents cross contamination of various samples used in experiments
[STANDARD SPECIFICATION]
-> Model : PURICUBE NEO 900 / PURICUBE NEO 1200 / PURICUBE NEO 1500 / PURICUBE NEO 1800
-> Inner(W x D x H)mm : 900x576x649 / 1200x576x679 / 1500x576x679 / 1800x576x663
-> Out(W x D x H)mm : 970x783x2100 / 1270x783x2100 / 1570x783x2100 / 1870x783x2100
-> Type : Class Ⅱ A2 (70% recirculation as downflow & 30% exhaust as outflow)
-> Filter : H-14 HEPA filter : 99.995% at 0.3㎛ / ISO class 5
-> Average Airflow Velocity Inflow : 0.40 m/s / 0.40 m/s / 0.40 m/s / 0.40 m/s
-> Average Airflow Velocity Downflow : 0.30 m/s / 0.30 m/s / 0.30 m/s / 0.30 m/s
[MAIN FEATURE]
-> Ergonomic design to give the increased convenience
- With a 10 degrees sloped design, it is possible to reduce tension in the arms, neck, and shoulders during the experiment. - Thanks to higher door glass, extensive visibility can be achieved. Since there is no frame at the bottom of the door glass, the visibility is not disturbed during experiment. - Due to the armrest, it is convenient to perform more than a certain amount of time, and the elbow can be fixed in the event of precise operation
- ① ∠10° Angle in front
- ② Frame-less window
-> Strengthen validation support using higher performance filters
- HEPA filter that enables to remove 99.995% of particles that have a size greater than or equal to 0.3㎛ - Optional ULPA filter that enables to remove 99.9995% of particles that have a size greater than or equal to 0.12㎛
-> Equipment that enables safe UV sterilizer
- UV sterilization will effect micro organisms at the surface of work table. - While UV light is on, front reinforcement glass door protects the user by blocking UV rays. - Automatically turns off UV light when opening the door or turning on the fluoresent lamp.
-> Checking the life and performance of filters through a differential pressure gauge
Measure the pressure of the bottom HEPA filter to check the performance and lifespan of the filter in real time, allowing more safe and efective testing.