BPA700 NIBP Simulator & Analyzer is a performance testing device that can be used standalone or connected to a PC. It contains specification range, adjustable pulse envelope to set precise values. Compatible with all types of blood pressure monitors, benefits R&D and compliance testing. Compliant test for IEC 80601-2-30, YY9706.230 and comply with simulator standard ISO/TS 81060-5.
OVERVIEW Compliant with IEC80601-2-30 standard testing, ensuring consistency and repeatability Adjustable blood pressure envelope to simulate different blood pressure Compatible with all types of blood pressure monitors, including wrist, arm, and tunnel Specification range: dynamic blood pressure range 10-300 mmHg and static pressure range 20-400 mmHg Dynamic pressure repeatability within 2mmHg at maximal pulse volume Provide Software Development Kit (SDK) to develop customized or automatic testing software Auto-sequence assists users in easily creating semi-automatic test procedures for more efficient and automated testing Optional Standard Assistant Software simplifies the medical standard with test sequences, options, parameters and pass criterions Standard Assistant Software with test report to help users keep records as HTML file Waveform Playback Software to play recorded raw data or customized waveforms
SPECIFICATION Parameters Specifications
Pressure Unit mmHg, kPa Mano
meter Range: 0 to 400mmHg
Resolution: 0.1mmHg Accuracy: ± (0.3% of reading + 0.5mmHg)
Pressure Source Target Pressure: 20 to 400mmHg Settle time: 5seconds Resolution: 1mmHg Accuracy: ± 0.5mmHg
Pulse Rate Range: 30 to 300bpm Accuracy: ± 1bpm Pulse Amplitude: 2mmHg max (500ml air réservoir) Pulse Volume: 0 to 2ml
Waveform Oscillometric Leak Test
Leak Test Time: 0 to 300seconds Target Pressure: 20 to 400mmHg Range: 0 to 300mmHg/min Overpressure Test Release Time: 0 to 300seconds Range: Automatic inflation Instantaneous: 0 to 400mmHg BP Envelope Shift Systolic Range: ± 20mmHg Diastolic Range: ± 20mmHg
BP Dynamic Range Diastolic Range: 10 to 250mmHg Systolic Range: 25 to 300mmHg
Dynamic NIBP Simulation Repeatability ±2mmHg 0.05mmHg (Standard deviation)
Self-Leakage <1mmHg/min (500ml air réservoir volume)
Environment Operating Temperature: 10℃ to 40℃ Storage Temperature: 0℃ to 50℃ Humidity: 0–90% RH (Non-condensing)
Shape Dimensions: 326.4 x 315 x 88mm Display: Graphic LCD Resolution 320 x 240pixels Weight: 4.5kg
Communication USB
Power AC 100-240V,50/60Hz