Intelligent operation system; Easy operation with visual and audio alarms; Multi-purpose service/maintenance interface; Profiling: sodium concentration and UF curve;
Main performance of this device:
Self-checking function; Carbonate dialysis& acetate; Double needle dialysis; Liquid level detector; Bubble detector; Blood leakage detector; Temperature and electric conductivity monitoring; Arterial pressure, venous pressure and transmembrane pressure monitoring; Rolling blood pump; Heparin pump; Dehydration amount controlled by capacity; Automatic disinfection cleaning program; Stand-by power of blood pump in case of power failure; Information display function of display screen.
*This device is designed and manufactured in strict accordance with the
requirements of national laws and regulations. The reference standards of design and manufacture are following: GB9706.1-2007; GB9706.2-2003; and YY0054-2003.
Technical Parameter
Size&Weight:
Size: 380mmx400x1380mm (L*W*H);
Area: 500*520mm;
Weight: 88KG;
Power supply:
AC220V,50Hz/60Hz,10A
Input power:1500W
Back-up battery: 30 minutes
Working Condition:
Water input pressure: 0.1Mpa~0.6Mpa, 15P.S.I~60P.S.I
Water input temperature: 5℃~30℃
Working environment temperature: 10℃~30℃ at relative humidity ≦70%
UF rate:
Flow range: 0ml/h~4000ml/h
Resolution ratio: 1ml
Precision: ±30ml/h
Blood pump&Spare pump:
Blood pump flow range: 20ml/min~400ml/min(diameter:6mm)
Spare pump flow range: 30ml/min~600ml/min(diameter 8mm)
Resolution ratio: 0.1ml
Precision: ±10ml or 10% of reading
Heparin pump:
Syringe size: 20, 30, 50ml
Flow range: 0ml/h~10ml/h
Resolution ratio: 0.1ml
Precision: ±5%
Monitoring system&Alarm setup:
Venous pressure: -180mmHg ~ +600mmHg, ±10mmHg
Arterial pressure: -380mmHg ~ +400mmHg, ±10mmHg
TMP pressure: -180mmHg ~ +600mmHg, ±20mmHg
Dialysate temperature: preset range 34.0℃~39.0℃
Blood leak detection: Photochromic alarm when erythrocyte specific volume is 0.32±0.02 or blood leak volume is equal or more than 1ml per liter of dialysate.
Bubble detection: Ultrasonic, Alarm when a single air bubble volume is more than 200μl at 200ml/min blood flow
Conductivity: Acoustic-optic
Disinfection/Sanitize:
1.Heat disinfection:
Time:30 minutes; Temperature: about 80℃, at flow rate 500ml/min;
Chemical disinfection:
Time: 30minutes, Temperature: about 36℃~50℃, at flow rate 500ml/min;
Chemical disinfection with heat:
Time: 45minutes, Temperature: about 36℃~80℃, at flow rate 50ml/min;
Rinse:
Time:10minutes, Temperature: about 37℃, at flow rate 800ml/min;
Storage Environment:
Storage temperature should be between 5℃~40℃, at relative humidity ≦80%
Packing:
Standard export packing (plywood box);