High Pressure Steam Sterilizer(250L)
FOB Price
Korea (North)
Main Products : sterilizer, autoclave, medical equipment, sterilization equipment
24 Annyeongnam-ro Hwaseong-si, Gyeonggi-do
High Pressure Steam Sterilizer(250L) Details
Brand Name
DELTACLAVE
Model Number
S-2510
Place of Origin
Korea
High Pressure Steam Sterilizer(250L) Introduce

A. Features

1. All functions are Automatic controlled by Micro-process, and the process are displayed on the board.

2. There is self-test function, checks conditions and Errors. Happened Error, beep and show Error No.

3. Air removal method is using powerful watering Vacuum pump.

4. Using the hinge Door and locking the door with radial type.

5. Display is the VFD or LCD TFT Touch screen.

B. Specifications

1 .Overall size: 755(W)×1730(H)×1470(L)mm or similar

2. Chamber size : 500(W)×500(H)×1000(L)mm or similar

3. Chamber type : Rectangular

4. Using pressure :1kg/cm2 ~ 2kg/cm2 or above

5. Using Temperature : 110℃ ~ 135℃

6. Timer : 0~99 min or similar

7. Chamber Material : All Stainless steel(316L)

8. Printer : Thermal Dot Matrix9. Automatic microprocessor control system including the following 6factory preset processes or equivalent

Process 1 Pre-vacuum wrapped

a) Pulsating pre-Vacuum

b) Steam sterilization 15 min at 132℃

c) Drying vacuum 20 min

Process 2 FLASH unwrapped

a) Pulsating pre-Vacuum

b) Steam sterilization 5min at 135℃

c) Dry vacuum 5min

Process 3 Gravity

a) Gravity air removal

b) Steam sterilization 20 min at 132℃

 
 

c) Dry vacuum 15min

Process 4 Gravity Liquid

a) Gravity air removal

b) Steam sterilization 15min at 121℃

c) Slow exhausted

Process 5 User select key

a) Select P1~P4 and then User key

b) Steam sterilization 3~ 60 min at 110℃~135℃

c) Dry vacuum 0~ 60min or Slow exhaust

Process TEST + P1= Vacuum Leak test

a) Pre-Vacuum

b) Vacuum isn’t down 1mmHg/min for 30 Min in test

c) No need test pack

 

Process TEST + P2= Bowie-Dick test

 
a) Pulsating pre-Vacuum

b) Steam sterilization 4min at 134℃ c) Dry vacuum 5min

d) Use the Test pack

10. Power Source : AC 380P, 3P, 3kW or 18kw (Steam generator built in)

11. External Steam : pressure 5kgf/Cm2, 99% saturated.

12. Safety device

(1) Safety valve: It is on the Chamber and Jacket that using above the 2.7Kgf/Cm2 and make the down the pressure.

(2) Door Safety device: When the chamber pressures is over the 0.3Kgf/Cm2, Door Safety device works and the door is not open.