Multi Parameter Patient monitor with Vital Signs ECG SPO2 NIBP Temp ambulance patient monitor ICU vital signs monitor
12.1" Portable Multi parameter Patient Monitor general ward and clinic monitoring bedside cardiac monitor Beta 12S
Portable Patient Monitor Quick Details:
Central StationTouch ScreenIoC/BIS (12.1 inch only)Suntech NIBP/Massimo SPO2Large 12.1 “ TFT LCD display7 Channel selectable waveformsProgrammable NIBP cyclesProgrammable alarmsSimultaneous display of all parameters
Hung-Kee Patient monitor Brochure 2022.01.pdf
Portable Patient Monitor Technical details:
Table ECG Specifications
Parameter | Specifications |
CMRR (common mode rejection ratio) | >90 dB |
HR measurement and alarm range | Adult/Pedi: 15bpm - 300bpm Neonate: 15bpm - 350bpm |
HR accuracy | 1bpm or 1%, whichever is greater |
ECG channel | Up to 7channels |
ST range | -2.0 mV - +2.0 mV |
ST accuracy | 0.02 mV or 10%, whichever is greater |
Pacemaker pulse rejection capability | Meets AAMI EC 13 |
ECG leads | 5-lead |
Alarm indication | Visual and audible indicators |
QRS indicator | Audible tone and heart symbol flashes |
ECG speed | 12.5 mm/s, 25.0mm/s, 50.0 mm/s |
ECG gain | ×0.25,×0.5,×1,×2,×4, Auto |
Bandwidth | Monitor mode: 0.5 Hz - 40 Hz Operation mode: 1.0 Hz - 25 Hz Diagnostic mode: 0.05 Hz - 100 Hz |
Tall T-wave rejection capability | Exceeds ANSI/AAMI EC 13 minimum recommended 1.2 mV T-wave amplitude |
Heart rate alarm response time | <12s |
Defibrillator protected | Tested with5kV |
Recovery time following defibrillation | < 5s |
Signal delay | 0.5s |
Auxiliary current (leads off detection) | Active electrode:< 100nA Reference electrode: < 900 nA |
Heart ratemeter accuracy and response to irregular rhythm | Ventricular bigeminy: 80 bpm±2bpm Slow alternating ventricular bigeminy: 60 bpm ±2bpm Rapid alternating ventricular bigeminy: 120 bpm ±2bpm Bidirectional systoles: 90 bpm±2bpm |
Response time of heart rate meter to change in heart rate | HR change from 80 bpm to 120 bpm: ≤ 8 s maximum HR change from 80 bpm to 40 bpm: ≤ 10 s maximum |
Time to alarm for tachycardia | Vent tachycardia (1 mVpp, 206 bpm): Gain 0.5, range 6.5 to 8.4s, average 7.2s Gain 1.0, range 6.1 to 6.9s, average 6.5s Gain 2.0, range 5.9 to 6.7s, average 6.3s Vent tachycardia (2 mVpp, 195 bpm): Gain 0.5, range 5.4 to 6.2s, average 5.8s Gain 1.0, range 5.7 to 6.5s, average 6.1s Gain 2.0, range 5.3 to 6.1s, average 5.7s |
ECG analog output | 0.67 Hz - 40 Hz |
Time to alarm for high heart rate | ≤10s |
Time to alarm for low heart rate | ≤10s |
( for more detailed technical data please refer to above PDF brochure)
Portable Patient Monitor Description:
The Beta 12S portable patient monitor is part of Hung-Kee’s premier patient monitoring solution designed to enhance patient care and support timely and accurate clinical assessment across general ward, daily checking and clinic environments. This portable patient monitor, power supplied by 100-240V~,50/60Hz, adopts 12.1'' color TFT LCD displaying real-time date and waveform. It can synchronously display seven-channel waveform and full monitoring parameters equipped with an optional 48mm thermal recorder. Maximum 64 units portable patient monitor could be connected to the central monitoring system via wire network to form a network monitoring system.
Portable Patient MonitorStandard Accessories:
5-lead ECG cable, SPO2 sensor, BP cuff, Temperature probe, Lithium battery, User manual, Power cord
Portable Patient MonitorOptional Accessories:
Aluminum trolley, wall mount stand
FAQ for Portable Patient Monitor:
1.Q: How long is the warranty time?
A: 24 Months
2. Q: When do you delivery?
A: Usually in 7-30 days depends on the order qty and available of some specific parts.
3. Q: Could you accept OEM order?
A: Yes, OEM and ODM are our main business.
4.Q: Do you work on SKD project?
A: Yes, we have more than 14 years experience on SKD service, full assembling procedure will be provided, and detailed guidance of local assembling work will be offered.
5. Q: Can I order a sample?
A: Sure, delivery sample by global express in charge.