Features:
l 12.1'' screen with true color, wide viewing angle, high brightness LCD display.
l HR/PR, NIBP, RR, ECG, SpO2, TEMP
l Simple and friendly operating display interface.
l Rechargeable lithium battery provides convenience for patients' moving.
l Playback and browse function for long term waveform and monitor data record.
l Optional printing output function。
l Auto double alarm with audible and visible signals
l Anti-defibrillation, anti-interference from high frequency electric knife
l A full-synchronistic lead multi-channel ECG display...
l This medical equipment can be used in ICU, CCU and operation room.
l Suitable for adult,pediatrics and neonate patients.
Standard configuration:
l HR/PR, NIBP, RR, ECG, SpO2, TEMP, Rechargeable lithium battery
Optional:
l Dual TEMP
l Dual IBP
l ETCO2
l Thermal recorder
l Wall mount
l Trolley
Specification:
ECG
Lead configuration
Standard 3-lead or 5-lead
3-lead RA、LA、LL,Lead method:I,II,III
5-lead RA、LA、LL、RL、V,Lead method:I,II,III,aVR,aVL,aVF,V
Increase
´250,´500,´1000,´2000
HR
Range
Adult 15 ~ 300bpm (beat/minute)
Neonatal baby/children 15 ~ 350 bpm(beat/minute)
Precision ±1% or ±1bpm,the larger prevails
Resolution 1 bpm(beat/minute)
Sensitivity
> 200 uV (Peak-to-peak value)
Input Impedance
> 5 (megohm)
Bandwidth
Diagnostic mode 0.05~130Hz
Monitoring Mode 0.5~40Hz
Operation mode 1~20Hz
Common Mode rejection Ratio
Diagnostic mode > 90 dB
Monitoring Mode > 100 dB
Operation mode > 100 dB
Pole Polarization Vvoltage Range
±300mV
Pacing Pulse Test
Test pacing pulse in accordance with the following conditions:
Amplitude: ±2 mV ~ ±700mV
Width: 0.1ms ~ 2ms
Risetime: 10us~100µs
Pacing pulse inhibition
When pacing analysis switch is on, pacing pulse in accordance with the following conditions are restrained, but affection against HR calculation.
Amplitude: ±2 mV ~ ±700mV
Width: 0.1ms ~ 2ms
Risetime: 10us ~ 100µs
Baseline Recovering Time
After defibrillation< 3 seconds
Signal Range
±8 mV (Peak-to-peak value)
Calibrating Signal
1mV(Peak-to-peak value),precision ±5%
ST Segment Measuring Volume
Measuring range: -2.0mV ~ +2.0mV
Measuring precision: Ranging -0.8 mV~+0.8mV, measuring error is ±0.02mV or ±10%,. The larger prevails.
No definition for other ranges.
RESP
Measuring Method
RA-LL impedance
RESP Impedance Measuring Range
0.3~3Ω
Base Impedance Range
200~4000Ω
Bandwidth
0.1~2.5Hz
RESP Rate
Range
Adult 0~120BrPM
Children and neonatal baby 0~150 BrPM
Resolution 1 BrPM
Precision ±2 BrPM
Asphyxia Alarm
10~40 seconds
NIBP
Measuring Method
Pulse wave oscillometry
Work Mode
Manual/Automatic/STAT
Measuring Interval of Automatic Measuring Mode
1,2,3,4,5,10,15,30,60,90,120,180,240,480 minute(s)
Measuring Time of STAT Mode
5 minutes
PR range
40 – 240 bpm
M