model specification |
WZS-50F6 | WZ-50C66T |
Flow rate(ml/h) | 10ml syringe:0.1-300 20ml syringe:0.1-399.9 30ml syringe:0.1-600 50ml syringe:0.1-1200 |
10ml syringe:0.01-300 20ml syringe:0.01-399.9 30ml syringe:0.01-600 50ml syringe:0.01-1200 |
Occlusion Pressure | WZS-50F6:40N,60N(±10N) WZ50C66T: L:300±100mmHg(40.7 ± 13.3 KPa); M:500±100mmHg(66.7 ± 13.3 KPa) H:800±200mmHg(106.7 ± 26.7 KPa) |
|
Bolus rate(ml/h) | 1200 (50ml syringe) 600 (30ml syringe ) 399 (20ml syringe ) 300 (10ml syringe ) |
1200 (50ml syringe) 600 (30ml syringe ) 399.9 (20ml syringe ) 300 (10ml syringe ) |
Total volume indication(ml) | 0.1-9999 | 0.01-9999 |
Volumetric accuracy | ≤2%(drive accuracy ≤1%) | ≤±2%(machine perision ≤±1%) |
Alarm | Near empty,Empty,Syringe dislocated,Syringe plunger disengaged,Malfunction,Forget to operate after power on,Occlusion,Low battery,Battery volume exhausted,AC power failure,Preset limited volume is finished | |
Dimension(mm) | 345(L)×186(W) ×128(H) | 307(L)×135(W) ×128(H) |
Net weigth(Kg) | 3.2 | 2.2 |
Power supply | AC230V/110V,50/60HZ(European Union),DC 12V The pump can run continuous over 4 hours using internal battery After it is charged for 16 hours at the rate of 5ml/h |