Ultraslim Portable Patient Monitor with 8 Inch Screen and Six Parameters Details
Brand Name
BENEMED
Place of Origin
China
Supply Type
OEM Service
Condition
New
Color
White
Ultraslim Portable Patient Monitor with 8 Inch Screen and Six Parameters Introduce

Specification

BenePM-8

- Unique slim design, stype with space-saving

- Parameters: ECG, SPO2, NIBP, HR, TEMP, RESP

- Networkable with Central Monitoring System

- 720 hours trend data analysis

- Compact design is convenient for mobile monitoring

- Built-in rechargeable battery with Max 6 hour operating time

- Optional:Touch screen, central monitoring(LAN/WLAN), Handle

ECG

Connection mode: 5 Leads(R,L,F,N,C or RA,LA,LL,RL,V)

Lead selection: I, II, III, AVR,AVL,AVF,V

Lead Mode: 3 Leads (R,L,F or RA,LA,LL)

Lead selection: I, II, III

Waveform: 1ch

Gain: x250, x500, x1000, x2000

HR and Alarm

Range: Adult: 15~300bpm

Neo/Ped: 15~350bpm

Accuracy: ±1% or ±1bpm, which great

Resolution: 1bpm

Sensitivity: >200(uVp-p)

Differential input impedance:≥5MΩ

ST segment Monitoring Range

Measure and Alarm:-2.0~+2.0mV

Measure accuracy: -0.8mV~+0.8mV

Heart rate

Measuring range: 0~350bpm

Accuracy: ±1bpm

Resolution: 1bpm

Sensitivity: >200UVp-p

Respiration

Method: Impedance between R-F(RA-LL)

Differential Input Impedance: >2.5MΩ

Measuring Impedance Range: 0.3~3.0 MΩ

Base line Impedance Range: 200~4000Ω

Bandwidth: 0.1~2.5Hz

Resp. Rate: Adult 2~120BrPM

Neo/Ped 2~150BrPM

Resolution: 1BrPM

Accuracy:+ 2BrPM

Allarm: 10~40S

NIBP

Method: Oscillometric

Mode: Manual, Auto

Measuring Interval in AUTO Mode: 1,2,3,4,5,10,15,30,60,90,120,180,240,

480(Min)

Measuring Period in STAT Mode: 5 Min

Pulse Rate Range: 40~240bpm

Alarm Type: SYS, DIA, MEAN

Resolution

Pressure: 1mmHg

Accuracy: Maximum Mean error: 5mmHg

Maximum Standard deviation: 8mmHg

SPO2

Display: Number of SPO2,waveform of capacity,Pulse

Measuring range: 0~100%

Resolution: 1%

Accuracy: 70~100% (±2%)

Alarm system: Audio-video alarm

Pulse rate range: 0~300bpm

TEMP

Channel: 2

Measuring and Alarm Range: 0~50ºC

Resolution:0.1ºC

Accuracy: 0.1ºC

Actualization interval: about 1 Second.

Average Time Constant: <10 second.