PRODUCT DESCRIPTION
LC3200 HPLC Liquid chromatography High Performance Liquid Chromatograph
Product Description
High performance liquid chromatograph (HPLC) is an instrument based on the principle of high performance liquid chromatography, which is mainly used to analyze less-volatile and thermal-unstable organic compounds with high boiling point and large molecular weight. It consists of reservoir, pump, sample injector, chromatographic column, detector and recorder.
Product Performance
High Pressure Pump
Customized high-precision gradient methods that guarantee extremely high gradient repeatability.
Patented two-stage suspension technology high pressure flow pump, improve the service life of the system
Built-in degasser for both high-pressure pumps and sampler cleaning fluid degassing
Autosampler
Completely self-developed autosampler. The polar coordinate design and various injection methods meet the needs of different customers. Equipped with an auxiliary air pump, there is a cooling mode, and an intelligent lighting mode is optional for easy observation.
Column Oven
Heating, cooling, two-way air bath temperature control, gentle temperature control, uniform heating, high stability.
Provide 2-way 250mm chromatographic column + guard column installation position to meet the needs of most users.
UV Detector
High-precision reflective lens, 1800-line blazed grating, to ensure extremely low stray light and great linear range.
Workstation
With local deployment and network deployment functions, to meet the needs of users of different scales
Product Specfications
I. Physical Specifications
1.1 Power requirements: 220V,50HZ
1.2 Ambient temperature: 4-40℃
1.3 Ambient humidity: 20-85% relative humidity(non-condensing)
II. System features
2.1 The system is modular designing.
2.2 Dual separated pump heads with specially designed integrated check valve.
2.3 Dynamic compression compensation technology ensure the flow accuracy and stability.
2.4 Active seal washing.
2.5 UV detector has wavelength scanning and dual wavelength detection functions.
2.6 The high-pressure pump adopts two-stage suspension transmission technology, combined with the suspension plug rod design of imported pump head, so as to prolong the service life of sealing ring and reduce the user's cost.
2.7 The proportional valve precisely controlled by the FPGA programmable logic ensure the accuracy of gradient proportion.
III. Performance Specifications
3.1 system
3.1.1 Qualitative repeatability(Retention time): RSD≤0.3%
3.1.2 Quantitative repeatability(Area): RSD≤0.6%
3.2 Pump
Flow rate range | 0.0001-10.000mL/min( increments :0.0001 mL/min) |
Max operating Pressure | 42MPa |
Pressure pulsation | <0.1MPa( Test condition: 10MPa , at 1.0ml/min methanol) |
Flow accuracy | ±0.2%( Test condition: 10MPa , at 1.0ml/min methanol) |
Flow precision | RSD≤0.06%( Test condition: 10MPa , at 1.0ml/min methanol) |
Gradient accuracy | ±1% |
Seal wash | With programmable on-line seal wash module |
3.3 Column oven
Temperature setting range | 5 °C ~ 85 °C. |
Temperature stability | ≤0.2°C/h |
Temperature controlling precision | ±0.1 °C |
3.4 Auto sampler
Injection Volume Precision | <0.25%RSD( full loop) ; <0.5%RSD( partial loop) ; <1%RSD( no waste) |
Linearity | >0.999 |
Cross Contamination(Carry Over) | <0.01% |
Sample Capacity | 120 samples( 1.5ml vials) |
Syringe | 250 μL( Standard, with 500μL,1000μL, 2500μL optionals) |
Sample Loop | 20μL( Standard, with 50μL, 100μL, 200μL optional) |
Optional sample chamber cooler | 4 °C ~ 30 °C(4 °C lowest) |
Injection volume range | 1-100 μL |
Injection mode | Full loop injection,partial loop injection,no waste injection |
3.5 Chromatography workstation
3.5.1 21 CFR Part 11 compliance and electronic registration of data capabilities. Meet regulatory requirements of data management integrity, security and audit trails
3.5.3 With Powerful functions.Support sample sequence management, analysis and data collection, processing, report, data management, three-dimensional spectra data collection, spectrogram retrieval, and other functions.
3.5.4 Friendly GUI design. Easy to use.
3.5.5 Stable and efficient database system, effective protection of user data security
3.6 DAD Detector
Noise | 1.5×10-5Au |
Drift | ≤1×10-3Au/h |
LOD | 5.0×10-9g/mL (naphthalene/methanol) |
Wavelength range | 190-800n |
Wavelength accuracy | ≤±1 nm |
Wavelength reproducibility | ≤±0.1 nm |
Linearity | ≥1.0×104 |
Extend functions | Wavelength bunching etc. |
3.7 FLD Detector
Noise | 1×10-5 Fu |
Drift | ≤2×10-4Fu/30min |
LOD | 5.0×10-10g/mL (naphthalene/methanol) |
Wavelength range | 200-650nm |
Wavelength accuracy | ±2 nm |
Wavelength reproducibility | ±0.2 nm |
Linearity | ≥1.0×104 |