X ray shielding lead glass
1.Technical details:
Type | Index of refraction | Density | Lead equivalent | light transmittance | PBo% |
ZF3 | 1.71720 | 4.46 | 0.24mmpb/mm | >85% | 61.03% |
ZF7 | 1.0600 | 5.2 | 0.33mmpb/mm | Less than 85% | 71% |
ZF6 | 1.75500 | 4.77 | 0.27mmpb/mm | Less than 85% | 65.06% |
2.Application:
a. Healthcare use of lead glass x-ray observation equipment, electron beam/ plasma generators and x-ray
TV detectors .Lead glass protects doctor and staff from x-ray irradiation with no glass discoloration or
deterioration in viewing quality.
b. Industrial use of lead glass
c. protect people from airport luggage inspection equipment in airports, and from the radiation testing
equipment or radioactive industrial products at industrial sites.
3. Manual installation of Lead Glass :
a. Lead glass is fragile, pay more attention when you delivery or remove the glass.
b. Forbidden to use the wet wooden frames, as the wood dry process can damage the lead glass.
c. The glass surface is forbidden to scrathe by the hard and sharg things.
d. Forbidden to scrub with the water to avoid mildew stain. If there is mildew stain can be removed by
absolute alcohol cotton ball.
e. Forbidden to touch with paint, thinner and other chemicals, to avoid oxygenizing the lead glass.
4. Process of Lead Glass :
We can produce three types of lead glass as follows:
a. Normal lead glass with different size- round,rectangular,square,irregular shapes.
b. Laminated glass-combined with normal tempered glass or clear float glass and lead glass.
c.Insulated glass-x ray shielding lead glass combined with normal tempered glass.